Calibration Calculator
Calculate error, deadband, hysteresis, and linearity from as-found and as-left readings.
For reference only — always follow your site calibration procedures and relevant Australian standards. This tool does not replace a certified calibration report.
Instrument Setup
Span: 100 | Tolerance band: ±1
0%, 25%, 50%, 75%, 100% of span
As-Found Readings
| Test Point | Expected | As-Found Up | As-Found Down | Error Up (%) | Error Down (%) | Hysteresis (%) |
|---|---|---|---|---|---|---|
| 0% | 0 | — | — | — | ||
| 25% | 25 | — | — | — | ||
| 50% | 50 | — | — | — | ||
| 75% | 75 | — | — | — | ||
| 100% | 100 | — | — | — |
Definitions
- Deadband:
- The maximum hysteresis across all test points — the largest difference between up-scale and down-scale readings at the same set point.
- Hysteresis:
- The difference between the up-scale (increasing) and down-scale (decreasing) reading at a given test point, expressed as a percentage of span.
- Linearity:
- Maximum deviation of the up-scale error curve from a best-fit straight line, expressed as a percentage of span.